Operation
Multi 3430
47
ba75865e19
04/2015
Note
For pH IDS sensors, you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 4.7.10).
Preparatory activities
Perform the following preparatory activities when you want to calibrate:
4.7.5
Calibration interval
The calibration evaluation is displayed as a sensor symbol.
After the QSC function has been enabled the sensor symbol is
replaced by the QSC color scale (see section 4.7.10).
After the specified calibration interval has expired, the sensor symbol
or the QSC color scale flashes. It is still possible to measure.
Note
To ensure the high measuring accuracy of the measuring system, cal-
ibrate after the calibration interval has expired.
Setting the calibration
interval
The calibration interval is set to 7 days (d7) in the factory.
You can change the interval (1 ... 999 days):
Eliminate the error according to chapter
6 W
Display
Calibration
record
Zero point
[mV]
Slope
[mV/pH]
1
Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.
2
Keep the buffer solutions ready.
1
Open the menu for measurement settings with
<MENU/
ENTER>
.
2
In the
Calibration
/
Calibration interval
menu, set the calibration
interval with
<
><
>
.
3
Confirm the setting with
<MENU/ENTER>
.
4
Quit the menu with
<M>
.
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......