01/2024
5.3
Settings
Making the set-
tings
Using
<S>
, switch from the measured value display to the main menu of the
settings. Then navigate to the setting menu of the sensor. The procedure is
described in detail in the IQ S
ENSOR
N
ET
system operating manual.
When operating the sensor in a flow through armature:
Remove the sensor from the flow armature.
Step 2:
Check:
1.
Submerse the sensor in the test solution.
2.
Compare the measured value to the nominal value.
Step 3:
Bring the sensor into the measuring position.
Once the check has been completed, the maintenance condition
has to be switched off manually. More detailed information on the
maintenance condition is given in the IQ S
ENSOR
N
ET
system
operating manual.
Setting
Selection/Values
Explanation
Measuring mode
U (mV)
Unit of the measured value in the measured
value display.
Temperature mode
°C
°F
Unit of the temperature measured value (Celsius,
Fahrenheit).
ORP shift
(only with measuring
mode mV)
-1250 mV ... +1250
mV
Here you can set the ORP zero point.
Temp. adjustment
-1.5 °C ... +1.5 °C
or
-2.7 °F ... +2.7 °F
The temperature compensation function enables
the temperature sensor to be balanced against a
reference temperature measurement.
Offset of the zero point:
±1.5 °C (±2.7 °F)
Displayed unit of the temperature:
see
Temperature mode
setting
Save and quit
-
The settings are stored.
The display switches to the next higher level.
Quit
-
The settings are not stored.
The display switches to the next higher level.