MIQ/IDS
Installation
13
ba77277d01
01/2024
3.6
Electrical connections: General instructions
Cable glands All SACIQ cables are fed from below through prepared openings in the housing
of the MIQ/IDS. Cable glands are provided with the MIQ/IDS to provide sealing
between the cable and housing as well as for strain relief.
General installation
instructions
Observe the following points when attaching connecting wires to the terminal
strip:
Shorten all the wires to be used to the length required for the installation
Always fit all the ends of the wires with wire end sleeves before connecting
them to the terminal strip
Any wires that are not used and project into the enclosure must be cut off as
closely as possible to the cable gland.
In each of the remaining free openings, screw in a small cable gland with
sealing ring and close them with a blind plug.
3.6.1 Power supply
Sealing ring 20 x 15 x 1 mm
Cable gland M16
Blind plug
WARNING
No free wires must be allowed to project into the enclo-
sure. Otherwise, there is a danger that areas safe to con-
tact could come into contact with dangerous voltages.
Always cut off any wires that are not in use as closely as
possible to the cable gland.
Component
Power supply
MIQ/IDS
Power is supplied via the IQ S
ENSOR
N
ET
system (see IQ S
ENSOR
N
ET
system oper
-
ating manual).