Sample preparation
Add 2 % ISA/K solution.
This sample conditioning solution creates optimum conditions for measuring.
It provides a constant ionic strength and similar diffusion potentials at the
reference electrode in standard solution and test sample.
Response times
The response time depends on the concentration range. It is
several seconds at high concentrations,
several minutes near the detection limit.
The measured value is stable if the value does not change by more than
0.1 mV within 30 seconds.
Interferences
Interfering ions: 10 % error with the following concentration ratio
(concentration ratio = interfering ion / measured ion):
The conditioning time depends on the previous history of
the ion sensitive electrode. To achieve a good responding
behavior, the concentration of the standard solution should
be similar to the concentration expected of the test sample.
During commissioning and after long periods of storage:
condition the electrode overnight.
Connect the electrode(s) to the meter and switch to the
measured value display (see operating manual of the
meter). The ion sensitive electrode is sufficiently condi-
tioned when the measured value no longer changes consid-
erably under otherwise constant conditions (temperature,
composition of the solution).
2
Remove any air bubbles in the bridge electrolyte by slightly knocking
against the shaft.
3
Calibrate according to the operating manual of the meter and the
analysis specification.
If you would like to have more detailed information concerning
sample preparation and measuring procedures, WTW provides
a large number of application reports for various applications.
Содержание wtw K 800 DIN
Страница 2: ...2 ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 5: ...BEDIENUNGSANLEITUNG ba15360d04 12 2021 K 800 DIN KALIUM IONENSELEKTIVE ELEKTRODE BEDIENUNGSANLEITUNG...
Страница 6: ...ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 13: ...K 800 Technische Daten ba15360d04 12 2021 13 Kabell nge Stecker DIN Stecker Anschlusskopf 16 mm 1 m...
Страница 14: ...14 ba15360d04 12 2021 Technische Daten K 800...
Страница 15: ...OPERATING MANUAL ba15360d04 12 2021 K 800 DIN POTASSIUM ION SENSITIVE ELECTRODE OPERATING MANUAL...
Страница 16: ...ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 23: ...MODE D EMPLOI ba15360f04 12 2021 K 800 DIN LECTRODE S LECTIVE D IONS DE POTASSIUM MODE D EMPLOI...
Страница 24: ...ba15360f04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 32: ...32 ba15360f04 12 2021 Caract ristiques techniques K 800...
Страница 33: ...MANUAL DE INSTRUCCIONES ba15360s04 12 2021 K 800 DIN ELECTRODO ION SELECTIVO PARA POTASIO MANUAL DE...
Страница 34: ...ba15360s04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 42: ...42 ba15360s04 12 2021 Especificaciones t cnicas K 800...
Страница 43: ......