Installation
DIQ/S 281
32
ba77244d01
08/2020
3.8.2
DIQ/S 28X-CR2/24V (24 V version)
Materials required
Wire end sleeves, suitable for the 24 V AC/DC feed line, with suitable crimp-
ing tool
1 x cable gland with sealing ring (included in scope of delivery of the DIQ/
S 28X).
Tools
Cable stripping knife
Wire stripper
Phillips screwdriver
Small screwdriver.
Preparing the 24 V
AC/DC line
WARNING
If the 24 V AC/DC supply is connected incorrectly, it may
represent a danger to life from electric shock. Pay atten-
tion to the following points during installation:
The DIQ/S 28X universal transmitter may only be con-
nected by a trained electrician.
The 24 V AC/DC supply must meet the specifications
quoted on the name plate and in chapter 9 T
(protective low voltage SELV).
The connection of the DIQ/S 28X universal transmitter
to the power supply may only be carried out when it is
not carrying any voltage.
When installed in a building, a switch or power switch
must be provided as an interrupt facility for the univer-
sal transmitter DIQ/S 28X.
The interrupt facility must
– be installed in the vicinity of the DIQ/S 28X universal
transmitter, easily accessible by the user, and
– be labeled as the interrupt facility for the DIQ/S 28X uni-
versal transmitter.
Rechargeable battery systems should have a deep discharge
protection. The DIQ/S 28X CR2/24V does not have any built-in
deep discharge protection.
1
Cut off the cable to the required length.
2
Strip the cable insulation for approx. 45 mm.
3
Bare the wires 1 and 2 and fit them with wire end sleeves.
Содержание WTW DIQ/S 281
Страница 2: ...DIQ S 28X 2 ba77244d01 08 2020 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 128: ...Messages DIQ S 281 128 ba77244e01 08 2020...
Страница 131: ......