SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
234
c) FlowTracker probes have been tested and have shown negligible flow
interference with the probe as much as 40-50
°
away from the preferred
alignment. At higher angles, the FlowTracker may see flow interference in the
sampling volume.
Figure 13:6 - Preferred Flow Direction to Avoid Flow Interference
Mounting Correction
13.11.3
Laboratory tow tank tests have indicated the FlowTracker2 probe and mount create flow
disturbance that may have a small impact (approximately 1.0%) on measured velocity
data. The mount is most commonly a top-setting wading rod using an S or J offset
bracket. Additional tests and modeling have looked at whether flowing water shows the
same flow disturbance effect seen in a tow tank (where the FlowTracker2 is towed
through a pool of stationary water).
By default, no correction for flow disturbance is applied to FlowTracker2 data. If you are
using a standard top-setting wading rod, and have decided to apply a correction,
SonTek typically recommends using a correction of 1.0%. For additional information
about the flow disturbance effect, contact SonTek directly.
To apply a flow disturbance correction, set the
Collection Settings menu.
No Correction
13.11.3.1
•
With the default setting of No Correction, the FlowTracker2 velocity data is used
directly without any correction for flow disturbance.
Custom Correction
13.11.3.2
•
This is selected when you want to apply a correction to account for flow
disturbance from the FlowTracker mount. With this option, you must specify the
value of the Mounting Correction within the range -5% to +5%.
•
When applied, the most common value for a standard top setting wading rod is
1.0%. A 1.0% correction means that measured velocity data is increased by 1.0%
to account for the effect of flow disturbance from the wading rod.
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...