7.4. Measurement Procedure
The basic outline of the
Measurement Procedure
required to perform a Discharge
measurement with a FlowTracker2 instrument is supplied in the following section. The
measurement procedure is based on wading principles, which is suited for low to
medium flow conditions depending on water velocity and depth.
a) Collect and verify Measurement Site Information before discharge measurement
b) Select the measurement site based on the criteria stipulated under
. Site and flow conditions does change over time and it is
recommend that existing measurement site is evaluated during each field
measurement exercise.
c) It is recommended for new measurement sites that the user perform spot checks
across the section at certain locations to get an estimation of the velocity range,
water depth and channel geometry.
d) Perform Pre Measurement Diagnostic tests in the region of the proposed
measurement section. Place the probe in moving water well away from any
underwater obstacles.
e) A graduated tagline or measuring tape should be spanned across the
measurement section to determine the station location. The graduated tagline or
measuring tape should be perpendicular to the main flow direction and channel
orientation.
f) Staff gauge readings with time should be taken at the start the discharge
measurement. If the discharge measurement is over extended period or at a
measurement site with rapid changing stage, intermediate readings should be
taken.
g) Enter Rated discharge (if applicable) of measurement site to calculate percentage
flow of each panel.
h) Divide the measurement section into a minimum of 23 stations (for sections >5m).
The number of stations required is dependent on the measurement section width
(see
ISO 748 – 2007, Hydrometry - Measurement of liquid flow in open channels
using current meters or floats
for recommendation of number of stations). The
flow per panel should not exceed 10% of the total flow, with 5% as the ideal
distribution.
i) The water depth at each station is measured using a top setting or universal rod.
A staff gauge board can also be used to determine the water depth at the location
of the station or sampling volume of the FlowTracker2.
j) The averaging time interval for data collection normally varies between 40 to 60
seconds for velocity measurements depending on flow conditions.
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...