88
Technical Specifications
1.3
1.3.1
Piston burette TITRONIC
®
500
Translation of the legally binding German version
(Release: 02. July 2020)
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1: 2013
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: 2010 for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin
: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
All common titration solutions.
As reagent water and all non-aggressive non-organic and organic fluids are allowed.
If using
combustible fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the
chemical industry.
For fluids with higher viscosity (
5 mm
2
/s), lower boiling point or affinity to outgas, the filling and dosage
speed can be adjusted.
Fluids with viscosity over 20mm
2
/s cannot be dosed.
Display:
3.5 inches -1/4 VGA TFT display with 320x240 pixels
Power supply
:
power supply 100-240 V; 50/60 Hz, power input: 30 VA
Use the power supply TZ 1853 only!
RS-232-C Interface
: separated galvanically through photocoupler, Daisy Chain function available
Data bits:
adjustable, 7 or
8
Bit (default: 8 Bit)
Stop bit:
adjustable,
1
or 2 Bit (default: 1 Bit)
Start bit:
static
1
Bit
Parity:
adjustable: even / odd /
none
Baud rate: adjustable: 1200, 2400,
4800
, 9600, 19200 (Default 4800 baud)
Address:
adjustable,
(0 to 15, default: 01)
RS-232-1
for computer, input Daisy Chain
RS-232-2
devices of SI Analytics
®
:
- Titrator TitroLine
®
7000 / 7500 / 7500 KF / 7750 / 7800
- Sample Changer TW alpha plus, TW 7400
- Piston burette TITRONIC
®
300 and 500, TITRONIC
®
110
plus,
TITRONIC
®
universal
,
-
Balances of the types Mettler, Sartorius, Kern, Ohaus, (for more, please contact us)
- Exit Daisy-Chain
USB Interface
:
2 x USB-type-A and 1 x USB-type-B
USB-type A
for connecting of USB keyboard, - printer, - manual controller, - data media
(e.g. USB stick) and USB-Hub
USB-type B
for connecting a PC
Ethernet Interface
: for connecting a local network (LAN)
Stirrer/pump
:
12V DC out, 500 mA
power supply for stirrer TM 235 and KF titration stand TM 235 KF
Housing
:
Material:
Polypropylene
Front keyboard: polyester coated
Dimensions:
15.3 x 45 x 29.6 cm (W x H x D), height incl. interchangeable unit
Weight:
approx. 2.3 kg for basic unit
approx. 3.5 kg for complete device incl. interchangeable unit (with empty reagent bottle)