Reset
Lab 955
28
ba77144e02
12/2017
7
Reset
You can erase the calibration values and reset (initialize) the measurement and
system settings.
7.1
Erasing the calibration values
This function serves to erase the last determined cell constant.
1.
Press
<On/Off
_
>
to open the menu to erase the calibration data.
Ini.C
is displayed.
2.
Use
<
><
> to display
no
or
YES
.
•
YES
: Erase the calibration values.
•
no
: Retain the calibration values.
3.
Confirm
with
<MENU/OK>.
The menu is finished. The meter switches to the measuring mode.
7.2
Resetting the measurement and system settings
1.
Switch on the meter with
<On/Off>
.
The display test appears briefly on the display.
2.
During the display test, press
<MODE>
to open the menu for the reset of
the meter settings.
Init
is displayed.
3.
Use
<
><
> to display
no
or
YES
.
•
YES
: Resetting the meter settings.
•
no
: Retain the meter settings.
4.
Confirm
with
<MENU/OK>.
The settings are reset. The menu is finished.
The meter switches to the measuring mode.
The calibration values are erased. All other meter settings
are retained.
The meter subsequently uses the last manually set cell con-
stant in the range 0.250 ... 2.500 cm
-1
.
The measuring system is not calibrated after a reset. Before
measuring, make sure the meter uses the cell constant suit-
able for the measuring cell.