Lab 865
pH value
ba77145e02
12/2017
23
5.1.2 Measuring the temperature
For reproducible pH and ORP measurements, it is essential to measure the
temperature of the test sample.
You have the following options to measure the temperature:
Automatic measurement of the temperature with the temperature sensor
(NTC 30 or Pt 1000) integrated in the sensor.
Measurement with an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes
0.1 °C
Automatic with tempera-
ture sensor
-
1 °C
Manual
1.
Measure the current temperature of the test sample.
2.
Set the temperature value with
<
><
>
.
or
In the
<F1>
/[Menu]/
Man. temperature
menu, set the temperature value
with
<
><
>
.