Setting Flows for the PFPD

 1.  Turn off the detector base heater, all 

detector gases, and the detector controller.

 2.  Allow the detector to cool and close the 

needle valve.

  3.  Install the barbed flow adapter in place of 

the flame arrestor.

 3.  Measure the column flow. It should be  

<2 mL/minute for optimum performance.

  4.  Open needle valve four turns.

Measure flows with a flow meter. DO NOT 

use the set point on the EPC.

  5.  Add 11.5 mL/minute H

2

. Be sure to add the 

column flow.

  6.  Add 10 mL/minute of Air 1.
  7.  Add 10 mL/minute of Air 2 (makeup).
  8.  Re-install the flame arrestor and turn on the 

detector heater.  Set it to 250 °C for sulfur 
or 300 °C for phosphorus.

 9.  When the detector temperature reaches  

100 °–150 °C, turn on the controller. Pulsing 
should start within a few seconds.

 10. Record the initial settings.

Tuning the PFPD

 

1.  Confirm installation of the blue, BG-12 filter or the clear, UV-34 filter.

  2.  Introduce a constant sulfur  source for tuning (e.g., a permeation device, PN 290411). 

OI Analytical recommends tuning the gas flows using a sulfur emission, regardless of 
the application.

  3.  After the sulfur equilibrates and the sulfur emission is on scale, click 

Hold 

in PulseView 

Monitor.

  4.  Adjust H

2

 flow up or down with approximately 1 mL/minute adjustments until the 

sulfur emission extends to 25 msec (if using a 3-mm combustor, the emission only 
extends to 18–20 msec). See the 

Model 5383 PFPD Operator’s Manual 

for more 

information.

  5.  Use Air 2 (makeup) to adjust frequency to between 2.5-4.0 Hz.  Re-adjust H

2

/Air 1 ratio 

as necessary.

  6.  Slowly close the needle valve by turning it clockwise until the PFPD goes into tick-tock 

mode.

  7.  Open the needle valve by turning it counterclockwise until the PFPD is out of tick-tock 

mode, then open another ½- to ¾-turn. Attaining tick-tock mode is sometimes not 
possible. In that case, just open the needle valve four turns.

  8.  Record the final settings.
  9.  For phosphorus applications, install the yellow RG-495 filter and increase Air 1 by 

1 mL/minute.

 10. Adjust the gate and board parameters for the chosen application.

Suggested Board Parameters

PMT voltage

  Set PMT Voltage to 500 V.  Monitor 

emission,  a d j u s t   P M T  v o l t a g e  
setting to adjust until the first 
sharp emission (at about 0.5 ms) 
t o   ~ 9 0 0 , 0 0 0   c o u n t s . Th e   P M T 
voltage can vary depending on the 
application.

Igniter current 

 2.8

Trigger level 

  75000 (initially), adjust to mid span.

For most sulfur applications:

*See the 

Model 5383 PFPD Operator’s Manual Appendix 

for instructions on how to properly set the zero offset 

and multiplier for each channel.

Suggested Gate Parameters

*  These settings are suggested starting points for adjusting the gates. Adjust-

ing the gate start time to eliminate the background HC emission may be 
necessary. This is especially important for phosphorus applications. Use 
the dual gate subtraction feature for matrices with high sulfur backgrounds 
(e.g., garlic, onion, etc.). See Chapter 5 in the 

Model 5383 PFPD Operator’s 

Manual

 and application note 1127 at www.oico.com for details.

  For more information contact OI Analytical Technical Support at  

(800) 336-1911 for assistance.

Element 

Gate 

Square Root

Sulfur 

6–24 msec 

On

Carbon 

1–3 msec 

Off

Phosphorus* 

4–15 msec 

Off

Parameter 

Channel 1 

Channel 2

Mode  

Sulfur Carbon

Interpolation   Spline 

Spline

Multiplier 32 

1

Output zero   

<100* 

<10,000*

Model 5383 PFPD Quick Reference Guide