Turbidity measurement (VisoTurb
®
900-P)
Multi 3620 IDS
68
ba77169d06
10/2021
Continuing with
three-point calibra-
tion
Repeat the steps 11 to 15 with the third standard solution. The new calibration
values are displayed after the last calibration step was completed.
9.2.5 Calibration data
Displays the calibra-
tion data
Calibration
The calibration protocol of the last calibration is available in the
menu
<MENU/ENTER>
/
Calibration record
. To open it in the measured value
display, press the
<CAL_>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in the
measured value display, press the
<MENU/ENTER>
key.
16.
If necessary, terminate the calibration as a two-point calibration with
<M>
.
The new calibration values are displayed.
or
Continue with three-point calibration.
15.09.2014 08:00
1010.0 FNU
Standard
3
Cal