Multi 3620 IDS
Dissolved oxygen
ba77169d06
10/2021
49
Freezing the mea-
sured value (HOLD
function)
With the HOLD function, you can freeze the current measured value. The dis-
played measured value stops changing until you switch the HOLD function off.
Stability control
(AutoRead )
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values. The display of the measured parameter flashes until
a stable measured value is available.
You can start a measurement with
Stability control
manually at any time, irre-
spective of the setting for automatic
Stability control
(see section 11.6.3) in the
System
menu.
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
If the HOLD function is active, you can, e.g. start a manual mea-
surement with stability control.
2.
Release the frozen measured value again with
<AR>
.
The HOLD function is switched off.
The [HOLD] status display disappears.
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
2.
Using
<MENU/ENTER>
, activate the
Stability control
function manu-
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the dis-
play of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the
Stability control
function manu-
ally with
<MENU/ENTER>
at any time. If the
Stability control
func-
tion is prematurely terminated, the current measurement data are
not output to the interface.
3.
Using
<MENU/ENTER>
, start a further measurement with
Stability
control
.
or
Release the frozen measured value again with
<AR>
.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Содержание Multi 3620 IDS
Страница 1: ...OPERATING MANUAL ba77169d06 10 2021 Multi 3620 IDS DIGITAL METER FOR IDS SENSORS 3620 IDS...
Страница 2: ...Multi 3620 IDS 2 ba77169d06 10 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3620 IDS Commissioning ba77169d06 10 2021 15 3 3 4 Setting the date and time 1 See section 4 5 5...
Страница 110: ...Glossary Multi 3620 IDS 110 ba77169d06 10 2021...
Страница 113: ......