Turbidity and Suspended Solids Transmitter
Turbidity calibration
A complete, normalized linearization curve of the probe is installed by the
manufacturer. No further calibration is possible (or required).
Suspended solids calibration
There are no calibration standards pre-defined for the system. Each meas-
urement must be calibrated individually, and it is possible to define two
suspended solids curves for use in different liquids.
Suspended solids calibration requires a sample to be evaluated in a
laboratory in order to set the relation between the measured point and the
actual value of the sample.
Averaging
The "Averaging" option sets the time period within which the measure-
ments are smoothed and averaged.
The current measured value is compared with the standard deviation of
the previous measured values. This feature ensures that a sudden devia-
tion in the liquid composition is eliminated in the measurement
Averaging
Units
mA output
Converter Setup
Quality improvement
No
Yes
Back Select
>
20 mA output
20 mA =
20.00 g/l
Back Select
OK
Units
NTU (Turbidity)
FTU (Turbidity)
FNU (Turbidity)
EBC (Turbidity)
% (Suspended solids)
mg/l (Suspended solids)
g/l (Suspended solids)
Kg/l (Suspended solids)
Back Select
OK
Fortsætter på side 5
Sensor Calibration
Averaging
Units
mA output
Cleaning
Solid curve
High Alarm
Low Alarm
Sensor error
Status
Factory settings
Service menu
Back Select
OK
4mA output
4 mA =
00.00 g/l
Back Select
OK
accordingly
Sensor Calibration
Yes
Use mA output
Yes
No
Back Select
OK
Averaging
0001 sec
(from 1 sek to 120 min)
Back Select
>
Yes
Adjust calibration
Select point to adjust
1 point Done
2 point Adjust
3 point
Back Select
OK
Calibration point 1
Type in lab. Value
10,2 g/l
Ok (/ Next)
Adjust
Back Select
Ok
Adjust
Calibration point 1
010,230 g/l
Back Select
>
Adjust
1
point
Adjust
Calibration point 1
0,00 g/l
Back Select
>
Adjust samples
Adjust
1. point
00010,2 g/l
Back Select
>
Adjust sample
1 Measured
10,2 g/l
2 Measured
8,11 g/l
3 Measured
5,87 g/l
Back Select
OK
Next
Select point
Adjust
2. point
00008,11 g/l
Back Select
>
Adjust
3. point
00005,87 g/l
Back Select
>
Sensor Calibration
Measurement samples
Adjust samples
Clear ”Measurement sample”
Clear ”Adjust samples”
Back Select
OK
1./2./3. point
Start sample?
Progress
[ 0% ]
Exit Start
Measurement samples
Select sample
1. point Measured
2. point Measured
3. point Measured
Back Select
OK
Clear ”Adjust samples”
Clear ”Adjustment sample”
No
Yes
Back Select
Ok
Start point
2. point
1. point
3. point
Next
Ok
Ok
Ok
Select point
Select point
1./2./3. point
Measureing
Progress
[||||||||||||||||| ] 45%
Cancel
Measure point 1/2/3
Done
Progress
[ 100% ]
Exit Next
Select point
Clear ”Measurement sample”
No
Yes
Back Select
Ok
Clear ”Measurement sample”
3 - Converter setup 1
Page3
1. Select "Yes" in the "Quality improvement" window.