pHix® Compact
EN 4.10 Version 2007
Maintenance / Cleaning of the electrode
32
Cleaning of the electrode
Use clean water for cleaning the electrode.
A 10% hydrochloric acid solution can be used for cleaning if the electrodes are
greasy. Protein coatings are removed with Pepsin cleaning fluid.
Electrode lifetime
The expected lifetime of the electrodes is dependent of the measurement. The
lifetime is given with reservation as the physical conditions as well as the
temperature, pressure and pollution of the fluid has great influence on the
stated lifetime.
The lifetime is based on
77° F / 25° C
, but if the temperature is doubled, the
lifetime is halved!
pH electrodes in purifying plants for ordinary household waste-water lasts
approximately one year, whereas the lifetime of pH electrodes in industrial
plants is dependent on measuring application and the process. Redox
electrodes have a lifetime of approximately two years, depending on cleaning.
Electrode spare parts
Item
no.
Description
160310 pHix® pH electrode, 0-point=4,6 pH
160311 pHix® pH electrode, 0-point=7,0 pH
160312 pHix® redox electrode
571030 Black rubber gasket for electrode
521409 Union ring w. Viton gasket for mounting of pHix® on pipe
- NOT
AVAILABLE IN THE US
521441 Viton gasket for union ring
- NOT AVAILABLE IN THE US
Содержание MJK pHix Compact
Страница 1: ...Manual EN 4 10 Version 2007 pHix Compact PH REDOX TRANSMITTER...
Страница 2: ...pHix Compact EN 4 10 Version 2007 2 Your notes...
Страница 5: ...pHix Compact EN 4 10 Version 2007 Declaration of conformity 5 Declaration of conformity...
Страница 23: ...pHix Compact EN 4 10 Version 2007 Mounting Connection examples 23 Connection to PLC_PLS...
Страница 35: ...pHix Compact EN 4 10 Version 2007 Dimensions 35 Dimensions Measurements are in mm inches US ONLY ALL OTHERS...
Страница 39: ......