pHix® Compact
EN 4.10 Version 2007
Buffer adjustment / Buffer adjustment using the tilt switch
28
In service (measurement). One
flash every 5. second.
Buffer adjustment in pH 4 buffer.
Buffer adjustment in pH 7 buffer. Buffer adjustment in pH 10 buffer.
Buffer adjustment using the tilt switch
Remember to clean the electrode as described earlier!
Only the built-in tilt switch is active when the switch ring has been removed for
more than 10 minutes.
The tilt switch inside pHix® Compact detects when the electrode tip is turned
upwards. When the electrode tip has been turned upwards for more than 30
seconds, pHix® Compact will go into buffer adjustment mode.
Proceed as described:
Remove pHix® Compact from the process, rinse the electrode, and turn the