Buffer adjustment / Buffer adjustment in general
24
Buffer adjustment
Buffer adjustment in general
In order to ensure a precise and stable pH measurement, pHix® Compact must
be buffer adjusted with regular intervals. Buffer adjustment is a calibration,
which is carried out in practice by dipping pHix® Compact in solutions (buffers)
with known pH values. The calibration is made by using these known pH values
as a reference point.
pHix® Compact is buffer adjusted using two buffers with pH values of 4 and 7
(acidic range) or pH values of 7 and 10 (alcalic range). For measurements
around pH 7 (neutral), both buffers can be used.
The sequence of different pH buffer solutions is not important. pHix® Compact
automatically detects which buffer is used for adjusting the point of origin, and
which buffer is used for adjusting the slope (sensitivity).
The first reference point sets the correct zero point for the measurement curve,
and the second reference point sets the correct slope for the measurement
curve.
pHix® Compact is, as standard, delivered to use buffer values at 4, 7 and 10,
but other values can be chosen via HART® communication.
First buffer adjustment with pH 7 as
reference.
Second buffer adjustment with pH 4
as reference.
Содержание MJK pHix Compact
Страница 1: ...Manual EN 4 10 Version 2007 pHix Compact PH REDOX TRANSMITTER...
Страница 2: ...pHix Compact EN 4 10 Version 2007 2 Your notes...
Страница 5: ...pHix Compact EN 4 10 Version 2007 Declaration of conformity 5 Declaration of conformity...
Страница 23: ...pHix Compact EN 4 10 Version 2007 Mounting Connection examples 23 Connection to PLC_PLS...
Страница 35: ...pHix Compact EN 4 10 Version 2007 Dimensions 35 Dimensions Measurements are in mm inches US ONLY ALL OTHERS...
Страница 39: ......