pHix® Compact
EN 4.10 Version 2007
Mounting / Eletrical mounting
19
US ONLY
ALL OTHERS
pHix® Compact is mounted correctly when the electrode tip is emerged min.
½" in / 10 mm
under the pipes inner pipe wall.
Eletrical mounting
Electrical mounting in general
pHix® Compact must not be installed in explosion hazardous locations!
pHix® Compact should be connected to an active 4-20 mA input or supplied
from a separate 12-30 VDC voltage source.
Max power consumption is 50 mW.
pHix® Compact 4-20 mA output(s) is galvanically separated from the liquids
potential.
Signal cable
pHix® Compact comes with either 1 or 2 galvanically separated signal outputs.
pHix® Compact with 1 x 4-20 mA output for pH or redox measurement.