Checking the zero point of the sensor
3.4
There are two methods to check the zero point of the sensor:
Measurement in a nitrogen atmosphere (recommended method).
Measurement in a sodium sulfite solution according to DIN EN 25814/ ISO 5814.
Test criterion
The sensor is OK if the measuring instrument displays < 0.5 % D.O.
saturation after 15 minutes.
4
What to do if ...
Error symptom
Cause
Remedy
No temperature value or
D.O. value
No connection between meter and
D.O. sensor
Establish connection between
meter and D.O. sensor
Cable defective
Return the D.O. sensor
Measured value too high
or too low or error
message:
Error
Coating on sensor cap
Clean the outside of the sensor
(
chapter 3.3.1)
Membrane damaged
Exchange the sensor cap
Service life of the sensor cap over
Check the sensor
Replace the sensor cap as
necessary (
Chapter 3.2)
Dirt inside the sensor cap and
sensor head.
Protective hood not properly
tightened.
Sensor cap untight or defective.
Dismantle the sensor cap
Clean the insides of the sensor cap
and sensor head (
Chapter
3.3.2)
Mount the sensor cap correctly and
tighten the protective hood as far
as it will go (
Chapter 3.2)
Replace a defective sensor cap as
Incorrect temperature
display
The temperature sensor is not
immersed deep enough in the
measuring solution
Observe the minimum immersion
depth
Temperature sensor defective
Return the D.O. sensor
Error message:
no cap
No sensor cap put on
Clean the sensor head and sensor
cap
Sensor cap not recognized
Return the D.O. sensor
Sensor cap defective
Replace the sensor cap