Installation
DIQ/S 281
ba76226e02
03/2023
16
arresters.
3.5
Connecting system components
3.5.1 General information
The IQ S
ENSOR
N
ET
system components are connected to form a functioning
unit in the following ways:
Stacked mounting of MIQ modules
Up to three MIQ modules can be installed and mechanically connected with
one another to form a stack at a single location. The contacts on the front and
rear automatically create the electrical connection between the MIQ modules
in the stack (section 3.5.2).
Distributed mounting of MIQ modules
Individual MIQ modules or an assembled module stack are installed at vari-
ous locations (for example, a central control unit and an MIQ branching mod-
ule for connecting IQ sensors at the edge of the basin). The spatial distance
is bridged with the aid of the SNCIQ or SNCIQ/UG cable (section 3.5.3).
Connection of IQ sensors:
IQ sensors are connected to an MIQ module with the aid of the SACIQ cable.
This cable is available in different lengths. The MIQ module must have a free
SENSORNET connection on the terminal strip. On the sensor side, the con-
nection is made with a plug that is standard for all IQ sensors (section 3.5.4).
The connection of the system components is described in the fol-
lowing sections exclusively with examples of the connection of MIQ
modules.
These descriptions similarly apply to system components that are
not mentioned separately in the examples (e.g. universal transmitter
DIQ/S 281).
Distributed mounting is only allowed in a star configuration. The
system must nowhere be closed to form a ring. Only use the
assembly material provided for the installation.