DIQ/S 281
Outputs
125
ba76226e02
03/2023
6.7.3
Fixed current value
Function
The
Fixed current value
function can be used to test the operativeness of the
instruments connected to the outputs: output different current values to the out-
put and, while doing so, check the behavior of the connected instrument.
6.8
Checking the condition of the outputs
This function provides a simple overview of the conditions of all outputs of the
combination output module (see section 4.7).
For the relays, the displayed condition is
open
or
closed
.
For the current outputs, the current value present at the outputs is displayed.
6.9
Behavior of linked outputs
6.9.1
Behavior in case of error
For linked relay outputs or current outputs, you can determine the behavior of
the outputs in case of errors.
Depending on the use of the output, the behavior in case of errors is set in the
following menus:
Error events
The specified behavior occurs with the following events or conditions:
Setting
Selection/Values
Explanation
Current output
Fixed current
value
Using
Save and quit
, the nominal
amperage that was input as
Inom
is output.
Inom
0 ... 20 mA
The nominal amperage that is
output.
The settings for other functions in the
Current output
menu as, for
example,
PID controller
and
Recorder
are retained while the
Fixed
current value
is carried out.
Output
Menu
Frequency controller Frequency with error
Pulse-width output
Error pulse width
Recorder
Current with error
(see section 6.7.1)