August 2020 – TD 316 OPERATING MANUAL – Motus Wave Sensor 5729
Page 15
Table 1-3: 5789 Cable pin configuration
1.5
User accessible sensor properties
All configuration settings that determine the behaviour of the sensor are called properties and are stored in a persistent
memory block (flash). One property can contain several data elements of equal type (Boolean, character, integer etc.).
The different properties also have different access levels.
To read the value of a certain properties you need to send ASCII string starting with the command get and then followed
by the property name to the sensor, see example below.
To change the content of a property an ASCII string starting with set and then followed by the property name and new
value in brackets need to be sent to the sensor.
For more information about sensor connection and configuration see
or
Get Interval
//When sending this string to the sensor, it will then return the value stored in
this property.
Interval 5729 19 100.000
//Returned from sensor, where 5729 is the product number, 19 is the serial
number of the sensor and 100.000 is the value stored as interval in seconds.
To change the value you might send the following command:
Set Interval(300)
//
This will change the value for this property to 30 minutes
Save
//Always end with save to store setting in flash
The interval will now be changed to 300 seconds.
Signal name
Sensor plug
MCIL10F
9-p D-Sub
USB
TXD
1
2
RXD
2
3
CAN_H
3
CAN_L
4
CAN_GND / NCG
5
NCR
6
NCE
7
BOOT ENABLE
8
Read Switch
GND
9
5 + Read Switch
4 (Black)
VPWR / Pos. Supply
10
1 (red)
RTS short to CTS
7-8
DSR short to DTR
4-6