Turb 430 IR/T
Operation
41
ba75507d09
08/2021
The data are output in the following order:
4.10
Reset
You can reset (initialize) all system and measurement settings.
4.10.1 Resetting the system settings
With the
System
/
Rücksetzen
function, all resettable settings are reset.
Settings for
Trübung
System settings
1 Date of storing
2 Time of storing
3 Selected ID
4 Measured value or
Upper/lower measuring range limit
(only with measured value status, OFL/UFL)
5 Unit of the measured value
6 Measured value status
* VALID: Measured value valid
* INVALID: Measured value invalid
* UFL: Measured value below the lower measuring range limit
* OFL: Measured value above the upper measuring range limit
7 Calibration status: Calibration type that was selected for the
measurement
8 Calibration status: Date of calibration
9 Calibration status: Time of calibration
If the connected external printer does not print, please check
whether the same baud rate is set at the Turb 430 IR/T and print-
er.
System setting
Default settings
Sprache
English
Baudrate
4800 Baud