TriOxmatic
®
690
Maintenance, cleaning, disposal, and replacement
29
ba25314e05
09/2020
5.5
Checking the sensor for freedom from zero-current
The sensor is zero-current free. Therefore, calibration of the sensor in
oxygen-free solution is not provided.
However, it is possible to check the freedom from zero-current
according to DIN EN 25814/ISO 5814. This is advisable if there is any
suspicion of a malfunction.
Test solution
1 g/l aqueous sodium sulfite solution, Na
2
SO
3
(addition of 1 mg/l of a cobalt(II) salt accelerates the removal of oxygen
from the solution)
The sensor should be in operation for at least 1 hour before the
inspection.
Test procedure
8
Fill a new WP 600 membrane cap and screw it on (see section
5.3 C
9
Connect the sensor to the measuring transmitter again.
10
After approx. 60 minutes polarization time, the sensor is ready
for operation.
11
Recalibrate the sensor.
1
Clean the sensor (see section 5.2 C
2
On the measuring transmitter, select
Saturation
(unit %) and
switch to the measured value display.
3
Submerse the sensor in the test solution.
4
Leave the sensor in the test solution for an hour.
5
Read the measured value.