Alarms
A range, consisting of a minimum and a maximum value, can be set for each channel.
If the alarm is enabled, the moment the measurement exceeds the set range the instrument will signal an error.
•
Select the channel on which we want to set the alarm.
•
Press
to enable the alarm on the channel.
•
The cursor will move to the value 'Minimum Threshold: Set the minimum threshold using the keys
and
press
to confirm.
•
The cursor will move to the value 'Maximum threshold: set the maximum threshold using the keys
and
press
to confirm.
Once in measurement, should the channel measurement exceed the values entered, the instrument will signal the
alarm both visually and by an acoustic signal.
Note
: The blue-coloured measurement indicates that the temperature has fallen below the set minimum threshold.
The red-coloured measurement indicates that the temperature has risen above the maximum set
threshold.
•
Reset
Access this Setup to reset the thermometer to its factory settings.
Resetting will reset all values to the default values shown in the table above.
Note
: When the reset is complete, the instrument returns to measurement mode and the parameters are reset to
default settings.
Log Settings Menu
TEMP:
Measurement Settings
LOG:
Data logger settings
SETTINGS:
General Instrument Settings
The table below shows the Setup menu structure for the parameter, for each programme the options the user can
choose and the default value:
Description
Options
Factory Settings
Logger settings Manual - Automatic - Programmed Automatic
Memory
Erase memory
No - Yes
No
•
Logger Settings
Access this menu to select the data acquisition mode:
•
Manual
: The data is only acquired when the user presses the
button;
•
Automatic
: The data is acquired autonomously by the instrument at regular intervals, set in this menu.
The acquisition interval (1 second to 99 hours) and the departure delay (00:00:00 = no delay) will be
requested;
•
Scheduled
: allows you to enter start date and time, end date and time, and acquisition interval.
The instrument will autonomously save data starting at the start date and time and stopping at the end date and
time.
Automatic logger - Setting the Automatic Data Logger
First set the acquisition interval (hh:mm:ss) then the delay (hh:mm:ss). Use the arrows
to change the flashing
value. Press
to confirm and move to the next value.