UM Temp74 PT100 Vio EN rev.1 19.05.2023 

 

Reference probe 

Through this menu we can select the Reference Probe. In the right column of the Multiview screen, for each 

channel, the difference of the measurement from the reference channel will be displayed. 

 

 

 

 

 

 

 

 

 

Offset 

The Offset menu allows you to adjust the measurement of each individual channel independently or to align all 

measurements to the reference channel. 

 

Offset adjustment (single channel) 

 

Select which individual channel you wish to adjust (Probe 1-2-3-4-) or select Align All to align all readings to the 

reference channel. 

 

Choosing a single channel will open the screen allowing adjustment of the channel value in real time. 

At the top the real-time reading is shown, at the bottom the adjustment value. 

 

Using the buttons 

 change the 'Adjustment' value until the desired 'Current Temperature' is reached and 

press 

 to confirm. 

Note

: An adjustment of a maximum of ± 5.00 °C from the current temperature can be made. 

If you select a channel with no probe connected, the display will show the error message: "CONNECT 
THE PROBE", accompanied by the icon flashing 

. If the offset has never been changed or a reset 

has been carried out, the message ''Offset 0.00 °C'' will be shown for each channel.

 

 

 

Offset adjustment (align all) 

This function allows the measurements of all channels to be aligned to the reference channel. 

 

Select our reference channel from the Setup menu 

Temp Settings -> Reference Probe and press 

 to 

confirm. 

 

In the Offset Probes menu, select Align All. 

 

The measurements of all channels will be aligned to the measurement of the reference channel. 

 

 

Calibration data 

Accessing the Calibration Data menu will display the calibration data for each channel. 

In addition to the value of the adjustment, the date and time at which it was made are displayed. 

If no offset has been entered, 0.00 °C appears. The same screen is displayed each time it is switched on. 

 

 

Resolution 

The Resolution menu allows you to select how many decimal digits to show in measurement. 

By default, the instrument is set with the highest resolution (0.01 °C). 
Using the buttons

 select the desired resolution and press

 to confirm. 

 

 

Stability filter 

In order for a value reading to be considered true, it is advisable to wait for measurement stability, indicated by 
the icon  . Using the buttons 

 select the desired degree of stability and press

 to confirm. 

 

Low: with variations less than or equal to 3 hundredths of a degree, the stability icon remains fixed and 

indicates the stable value. 

 

Nor: with variations less than or equal to 2 hundredths of a degree, the stability icon remains fixed and 

indicates the stable value. 

 

High: with variations less than or equal to 1 hundredth of a degree, the stability icon remains fixed and 
indicates the stable value. 

+0,02

-0 ,05

+0,04

37,05 °C
37,07 °C
37,00 °C
37,09 °C

Multiview

Содержание Temp74 PT100 Vio

Страница 1: ...INSTRUCTION MANUAL Temp74 PT100...

Страница 2: ...tilingual copy Espa ol Para reducir los gastos este manual ha sido impreso exclusivamente en italiano e ingl s Contacta tu distribuidor para recibir la copia digital multi idioma Deutsch Um umweltfreu...

Страница 3: ...nstrumental Characteristics 6 Technical Data 6 Description Instrument 6 Display single channel 6 Multiview screen 7 Keyboard 7 Key Functions 8 LED 8 Installation 8 Supplied Components 8 Commissioning...

Страница 4: ...ngs 14 Date and time 14 Units of Measurement 14 Language 14 Auto off 14 Sound 14 Password 15 Temperature measurement 15 Measure 15 MAX MIN function 15 HOLD function 15 Other functions 15 Automatic swi...

Страница 5: ...ormation NOTE for useful product information Warning symbols Additional documents providing security information The following documents can provide the operator with additional information to work sa...

Страница 6: ...mended to occasionally clean the instrument casing with a damp cloth and mild detergent This operation should be carried out with the instrument switched off and only by experienced and authorised per...

Страница 7: ...MAX MIN reading Yes Offset Adjustment Yes for each channel independently Internal memory 200 000 data Data logger Manual Automatic Programmed Sampling 1 min 99 hours Working conditions Temperature 0 6...

Страница 8: ...X and MIN values recorded Press for 3 seconds to activate and deactivate the HOLD function and to deactivate the MAX MIN recording function if active Directional key DOWN RM to scroll through menus an...

Страница 9: ...he display brightness Only at this point do the buttons regain their function LED All instruments are equipped with a two colour LED red and green that provides the user with important information on...

Страница 10: ...carrying case Only use the original carrying case to transport the instrument Contact your local distributor if you need to buy it back The inside of the case is shaped to accommodate the instrument...

Страница 11: ...e below shows the Setup menu structure for the parameter for each programme the options the user can choose and the default value Description Options Factory Settings Sensor ID Probe 1 Probe 2 Probe 3...

Страница 12: ...annels to be aligned to the reference channel Select our reference channel from the Setup menu Temp Settings Reference Probe and press to confirm In the Offset Probes menu select Align All The measure...

Страница 13: ...the table above Note When the reset is complete the instrument returns to measurement mode and the parameters are reset to default settings Log Settings Menu TEMP Measurement Settings LOG Data logger...

Страница 14: ...pendently once the set date and time is reached and will stop Note If recording takes place while there is a single channel measurement on the display data concerning only that channel will be saved I...

Страница 15: ...er battery life Brightness Access this Setup menu to choose between three different display brightness levels Low Normal High Note Keeping the display at high brightness at all times negatively affect...

Страница 16: ...appears on the display Always wait for the measurement to become stable Note If no temperature probe is connected the message CONNECT A PROBE appears When connecting or disconnecting probes the instru...

Страница 17: ...corrected and the adjustment value with flashing cursor using the keys correct the value to Current Temperature to 0 0 C press to confirm and record the adjustment Note The adjustment affects the ent...

Страница 18: ...e manufacturer will repair or replace defective components This guarantee applies only to the electronic part and does not apply if the product has been damaged misused exposed to radiation or corrosi...

Страница 19: ...18 UM Temp74 PT100 Vio EN rev 1 19 05 2023 XS Instruments Via della Meccanica n 25 41012 Carpi MO ITALY Tel 39059 653274 Fax...

Отзывы: