10
UM Temp74 PT100 Vio EN rev.1 19.05.2023
Setup Menu
TEMP:
Measurement Settings
LOG:
Data logger settings
SETTINGS:
General Instrument Settings
•
In measurement mode, press the
button to enter SETUP mode.
•
Within the setup, move through the different menus using the
buttons and press the
button to access
the submenu you wish to edit.
•
Press
to return to the previous menu and to exit without saving.
Temp Settings Menu
The TEMP Settings menu contains all settings concerning the measurement.
TEMP:
Measurement Settings
LOG:
Data logger settings
SETTINGS:
General Instrument Settings
•
In measurement mode, press
to enter the SETUP menu.
•
Press the
button to access the TEMP SETTINGS menu.
•
Use
to select the programme you wish to access.
The table below shows the Setup menu structure for the parameter, for each programme the options the user can
choose and the default value:
Description
Options
Factory Settings
Sensor ID
Probe 1 - Probe 2 - Probe 3 - Probe 4
Reference probe Probe 1 - Probe 2 - Probe 3 - Probe 4
Offset
Probe 1 - Probe 2 - Probe 3 - Probe 4 - Align all
Calibration Data
Resolution
1 °C - 0.1 °C - 0.01 °C
0,01 °C
Stability filter
Low - Normal - High
High
Alarms
Probe 1 - Probe 2 - Probe 3 - Probe 4
Deactivated
Reset
No - Yes
No
•
Sensor ID
Access this menu to enter an ID (identification code) to one or more connected channels.
•
Select which channel you wish to identify.
•
Starting with the digit with the flashing cursor, enter the desired value using the keys
and press
to
move to the next digit.
•
A code with a maximum length of 10 digits can be entered.
Note
: If you select a channel with no probe connected, the display will show the error message: "CONNECT THE
PROBE", accompanied by the icon flashing
.
Содержание Temp74 PT100 Vio
Страница 1: ...INSTRUCTION MANUAL Temp74 PT100...