EN
Logger has been set. Press
to start recording; the icon
starts flashing, indicating that the
storage is in progress. The number next to the
M+
icon indicates how much data has been saved for
that parameter.
Press again
to end the recording.
Note
: automatic recording is suspended when the measurement parameter is changed.
•
Example of manual Data Logger mode
Example: recording a Conductivity value in manual mode
•
Access the
LOG SETTING P8.
0 setup menu.
•
Enter the
LOG TYPE P8.
1 menu, press key
and move with the directional keys to
MANUAL
.
•
Confirm with
and return to measure mode, go to the
COND
screen
.
Press
to save the value. The number next to the
M+
icon indicates how much data has been saved
for that parameter.
Note
: Manual or automatic saving of a value is confirmed by a sequence of flashes of the green LED
•
Recall Memory
•
In measure mode in the parameter of interest, press key
to enter the RECALL MEMORY mode.
The last saved data is shown on the display.
•
As indicated by the string
, use the directional keys to scroll through the different stored values.
The number next to the
M+
icon indicates the save slot.
•
Press
to return to measure mode.
•
Clear the saved data
•
To clear the data stored in the instrumental memory, access the
CLEAR DATA P8.2
setup menu and
select
YES
.
IMPORTANT
: Factory reset of the pH, ORP and Cond parameters does not delete the stored data
Instrument Setup menu
•
In measure mode, press key
to access the SETUP menu.
•
Use the directional keys to move to
SETTINGS P9.0
and access the menu by pressing the key
.
•
Move with the keys
and
to select the program to access.
The table below shows the setup menu structure for the general settings of the instrument; for each
program, there are the options that the user can choose and the default value: