2
2-11
Using Calibrator
T
O
CALIBRATE
THE
F
IERY
USING
THE
DTP41
1.
Start Calibrator.
For instructions on starting Calibrator, see
2.
Choose X-Rite DTP41 as the measurement method.
3.
In the Check Print Settings pane, choose the desired calibration set.
Choose the appropriate calibration set for the type of media you use most often.
N
OTE
:
For this calibration to take effect, the calibration set must be associated with one
or more output profiles. There is no need to make any new associations for the default
calibration set, since it is already associated with the default output profile.
4.
In the Generate Measurement Page pane, click Print.
The Print Options dialog box is displayed.
5.
Choose the appropriate options, and click Print.
For Page Type, choose either 21 or 34 Sorted Patches.
Paper Size is automatically set to LTR/A4 for 21 Sorted Patches, or 11
x
17/A3
for 34 Sorted Patches.
For Input Tray, choose the paper source for printing the measurement page.