Trigger out
The camera can generate digital output signals which are used as trigger output.
Digital output (trigger output) mode
External trigger out 1; External
trigger out 2
These triggers are available via the universal camera
connector on the backside of the camera (see section
4.1 for the pinout of the trigger cable).
Table 5-10 Digital output (trigger output) mode
Trigger out source
High
Constant high voltage
Low
Constant low voltage
Start of Frame
A pulse is generated at the start of a frame
Start of Line
A pulse is generated at the start of a line
Table 5-12 Tigger out source
It is possible to set both outputs with the same trigger out source.
Digital output (trigger output) voltages
Trigger
out
voltages
and
impedance
Low
≤ 0.4V*
High
≥ 2.9V*
R
OUT
= 100Ω
*R
LOAD
≥1kΩ
Table 5-11 Trigger out voltages and impedance