Local Lockout
M370078-01
3-37
3
Local Lockout
Local lockout is a feature that allows the front panel to be locked so that
accidental button presses are ignored. This feature is often used to lockout
the front panel when you are controlling the power supply from a remote
location. When in local lockout mode, the front panel will display
LOCL
Loc
whenever a button is pressed or a knob is turned.
Enabling Local Lockout
To enable local lockout:
1. Turn the 9-position mode control to the PGM position or press the
rotary Adjust/Enter control if the control knob is already at the PGM
position.
The display will show
rE
in the output voltage display and the
selected remote interface in the output current display.
2. Turn the rotary Adjust/Enter control until the display shows
LOCL
Loc
.
3. Press the rotary Adjust/Enter control.
4. Turn the rotary Adjust/Enter control until the display shows
LOCL On
.
5. Press the rotary Adjust/Enter control.
Local lockout has been enabled.
The SCPI command (s) for these instructions are:
[:]SYSTem:REMote:STATe
Disabling Local Lockout
To disable local lockout:
1. Turn the 9-position mode control to the PGM position or press the
rotary Adjust/Enter control if the control knob is already at the PGM
position.
The display will show
rE
in the output voltage display and the
selected remote interface in the output current display.
2. Turn the rotary Adjust/Enter control until the display shows
LOCL
Off
.
3. Press the rotary Adjust/Enter control.
Local lockout has been disabled.
Содержание Watt Series Programmable DC Power Supply XG...
Страница 2: ......
Страница 3: ...XG 850 Watt Series Programmable DC Power Supply Operating Manual...
Страница 6: ......
Страница 18: ...xvi...
Страница 30: ...1 8...
Страница 52: ...Local Operation 3 8 M370078 01 Figure 3 3 Front Panel Menu System...
Страница 82: ...Local Operation 3 38 M370078 01 The SCPI command s for these instructions are SYSTem channel s REMote STATe...
Страница 86: ...Local Operation 3 42 M370078 01 Figure 3 5 Split Supply Operation V V Common Power Supply Power Supply Load...
Страница 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370078 01...
Страница 198: ...5 72...
Страница 262: ...A 36...
Страница 284: ...IX 4...
Страница 285: ......