Calibrating the Device
If you own multiple devices make sure that the serial number on the front and back of the calibration plate
matches the serial number of your i1 Paint device. Refer to the user guide of your measurement software
for additional information on the calibration process.
1. In the center of the calibration plate is a slider to protect the white calibration tile. Open this
protective slider until it clicks into place to access the white reference ceramic tile on this calibration
plate.
2. Place the i1 Paint measurement device on the calibration plate. The device must seat firmly on the
calibration plate.
3. Start the calibration mode in the software, and press the measurement button on the device or in
the software. Status indicator light switches off during calibration. Once the device is successfully
calibrated the status indicator light pulsates white. If the calibration fails the status indicator light
pulsates red and then returns to solid white.
4. When the calibration is finished you should close the protective slider for the white reference
ceramic tile.
A clean white reference ceramic tile is essential for providing accuracy of your measurements. The
protection cover should always be closed when the white reference ceramic tile is not in use. If
necessary you can clean the white reference ceramic tile with isopropyl alcohol and a soft clean
cloth.