B A S F S P E C T R O P H O T O M E T E R 1 2 / 6
20
Calibrating the Instrument
1.
Select the calibration option from the Settings menu as previously explained (see Entering Settings Mode).
The message “Calibration Required” appears if the instrument needs to be calibrated. If calibration is not
currently needed, the time remaining before the next calibration appears. To exit the calibration mode
without calibrating, tap
Ignore
.
2.
Tap
Calibrate
to continue with the calibration.
3.
Position the instrument on the white tile as previously explained. Swipe the screen to the left for additional
information on positioning.
4.
When ready, press the top
Measure
button or tap the Calibrate icon
in the screen. Do not touch the
instrument throughout the measurement sequence.
NOTE:
If an error message appears after white calibration, try measuring the white tile again. If an error still
occurs, clean the white calibration tile as explained in the Appendices.
5.
After calibration is completed, remove the instrument from the white tile and position it over the effects tile as
previously explained. Press the top
Measure
button or tap the Calibrate icon
in the screen. Do not
touch the instrument throughout the measurement sequence.
6.
After reference verification is completed, remove the instrument from the calibration reference and return the
reference to its storage location.
Current calibration
status