REMOVABLE FOOT PLATE

Removable footplate for 

AMBULATION

.

51924-R01

Squeeze the yellow bar & 

push down/pull up to activate 

pneumatic stand assist. 

Step on yellow pedals to close 

and open legs to go around 

chairs & ambulate.

Adjustable seat pad 

configurations.

Pull yellow levers to adjust 

clamp. Turn to loosen and 

tighten adjustable shin pad.

800.255.3126 | wyeastmedical.com

VISTA

400

Instructions For Use

Содержание VISTA 400

Страница 1: ...wn pull up to activate pneumatic stand assist Step on yellow pedals to close and open legs to go around chairs ambulate Adjustable seat pad configurations Pull yellow levers to adjust clamp Turn to loosen and tighten adjustable shin pad 800 255 3126 wyeastmedical com VISTA400 Instructions For Use ...

Страница 2: ... Needed Position Patient s Feet in Footrest Position Shin Support Against Shins Secure clamp A Support Must Be Below Knee Joint Transport Patient Have Patient Grasp Handlebar and Stand Position Seats Under Patient and Have Patient Sit Adjust Shin Vertical Support Secure clamp B All Yellow Areas on Equipment Are Adjustable UNLOCK WHEELS Adjust Sitting Height ...

Страница 3: ...800 255 3126 wyeastmedical com 51928 R03 Vista Non Powered Stand Aid Operator s Manual Page 1 of 17 VISTA Operator s Manual OPERATOR S MANUAL VISTA NON POWERED STAND AID Model No 51871 ...

Страница 4: ...es the right to confirm the existence of any claimed defect and to determine whether any damage was caused by neglect misuse unauthorized alteration or failure to follow maintenance instructions With respect to a claim covered by this warranty the defective part will be replaced or repaired at no charge provided the defective part is returned to Wy East The user has the responsibility for installa...

Страница 5: ... Seat Position Out________________________________________________________ 8 Seat Position Up and Out__________________________________________________ 8 Seat Height Adjustment ___________________________________________________ 9 Legs Open Close Function ________________________________________________ 9 OPERATING THE VISTA NON POWERED STAND AID _____________________________ 10 TECHNICAL SPECIFIC...

Страница 6: ...njury Keep this manual with the Vista Non Powered Stand Aid and refer to it as required Contents of this manual are subject to change without prior notice to users DEFINITIONS The words WARNING CAUTION and NOTE carry specific meanings and should be carefully reviewed WARNING The personal safety of the patient or user may be involved Disregarding this information could result in injury to the patie...

Страница 7: ...ta has a compact footprint allowing for greater maneuverability in tight spaces Intended for use in acute care long term care outpatient care and home care It is not recommended for use in a shower Prior to using the Vista for transfer or any other procedure the patient must be assessed by a qualified professional to determine the suitability of using the Vista with that patient Patients should be...

Страница 8: ...sfer procedure 5 Ensure the patient is properly seated in the Vista before transporting Never transport a patient while they are standing in the Vista 6 Never exceed the safe working load SWL is 400 lbs 7 Do not power wash the Vista 8 When cleaning and disinfecting use a 5 25 chlorine bleach dilution and an EPA registered Hard Surface Disinfectant on painted and plastic surfaces On all other surfa...

Страница 9: ...ion the lever at an angle to allow clamping pull the yellow handle back see Fig 2a rotate to desired position see Fig 2b and release 3 You are now ready to rotate to clamp Shin Guard Adjustment Toward Away from Patient Adjust shin guard to ensure that shin guard rests against the patient s shins while patient is sitting with lower leg at 90 to thigh where the thigh is parallel with the floor Shin ...

Страница 10: ... of 17 VISTA Operator s Manual Seat Position Up To move seat up pull up and rotate seat as shown Seat Position Out To move seat out push up on yellow knuckle and rotate seat as shown Seat Position Up and Out To move seat up and out follow the instructions for the two previous features FEATURES ...

Страница 11: ... the yellow lever and pull up or push down on the handlebar NOTE Adding weight on the seat pads can aid in height adjustment Legs Open Close Function 1 Push down on the left yellow pedal to open legs 2 Push down on the right yellow pedal to close legs Allows clearance for the caregiver to get around a chair wheelchair or commode and properly place the feet and set the shin guard FEATURES ...

Страница 12: ...l their feet are completely supported by the footrest and their legs are in a position as close to 90 degrees as possible Fig 4 4 LOCK the rear WHEELS Figure 3 Figure 4 5 Move shin guard forward to rest against the patients shins Lock Cam Lever 6 Adjust the shin guard until it rests 2 or more inches below the patients knees patellae This prevents the patient s kneecaps being pushed upward when the...

Страница 13: ...he patient hold the handlebar This keeps their arms in a SAFE position for transfer 13 DO NOT transport the patient in a standing position 14 UNLOCK the rear casters and move the Vista to the location of intended transfer 15 Once the patient is positioned over the intended area of seating LOCK the rear WHEELS 16 Have the patient stand move the SEATS out and have the patient sit 17 Once the patient...

Страница 14: ...eter 46 1 Leg Under Bed Stretcher Clearance 3 4 Standard Door Width Clearance 23 7 Footplate Dimensions 14 x 17 x 14 Frame Steel Aluminum FDA Device Class Class 1 FDA Product Code FSA FDA Regulatory Number 880 5510 ENVIRONMENTAL CONDITIONS FOR SHIPMENT AND STORAGE FOR USE Temperature 14 F to 104 F 10 C to 40 C 41 F to 104 F 5 C to 40 C ambient Relative Humidity 20 to 80 non condensing 20 to 80 non...

Страница 15: ...Not recommended for use in shower Do not pressure wash the unit ONCE A MONTH Check for loose or missing nuts and bolts Clean unit wipe off dust TROUBLESHOOTING SYMPTOM POSSIBLE CAUSE ACTION Seat height adjustment does not release or lock in position selected Release lever is not functioning properly Adjust release lever pivots Gas spring release cable s is not functioning properly Adjust or replac...

Страница 16: ...e following information with your request 1 Serial and Model numbers of your Vista Non Powered Stand Aid 2 Purchase date if available 3 Part number and description of the replacement part needed see Replacement Parts section on page 15 of manual 4 Shipping and billing instructions including complete address RETURNING ITEMS To return parts or goods contact Wy East Medical for authorization Contact ...

Страница 17: ... 51675 51670 51899 51854 51647 51852 51671 51672 51687 51687 REPLACEMENT PARTS This manual provides maintenance and service information for the Vista Non Powered Stand Aid If replacement parts are required contact WY EAST MEDICAL with the serial number of your Vista Non Powered Stand Aid to ensure the accuracy of replacement parts ...

Страница 18: ...800 255 3126 wyeastmedical com 51928 R03 Vista Non Powered Stand Aid Operator s Manual Page 16 of 17 VISTA Operator s Manual NOTES ...

Страница 19: ... s Manual Page 17 of 17 VISTA Operator s Manual CONTACT INFORMATION For questions or information regarding other Wy East Medical products contact Wy East Medical Corporation 16700 SE 120th Ave Clackamas OR 97015 USA 503 657 3101 800 255 3126 wyeastmedical com info wyeastmed com ...