What to do if...
73
6
What to do if...
Flashing
1100
display
Measured values
that are obviously
too high
Error message
ERR
Cause
Remedy
– Measuring range
exceeded
– This cannot be remedied.
Sample is not suitable for
the measurement
Cause
Remedy
– Cuvette contaminated
– Clean the cuvette
– Cuvette scratched
– Replace the cuvette
– Cuvette misted up
– Adjust the temperature of
the sample before
measuring
– Air bubbles in the cuvette – Remove air bubbles
– Measuring instrument
not correctly calibrated
– Calibrate the instrument
Cause
Remedy
– Internal error
– Have the instrument
repaired by the WTW
service department