6.1.3
Testing the interfaces
You can send a test string "Test string RS232 (or RS485)" to the selec-
ted interface.
Sending a string
Example Test string
Fig. 6-1
RS485 test string
1
Measuring mode
Call up the
Service
menu: Simultaneously press the keys,
4
and
5
. The request
Enter PIN
appears
.
2
Input the PIN digit-by-digit with
68
. Change the digit
with
57
and confirm with
3
. The
Service Mode
menu appears
.
3
Select the
RS232
or
RS485
menu item with
68
.
4
Confirm the selection with
3
. The test string is sent to the
selected interface. A confirmation appears on the display.
------------------------------------------------------------
RS485 test string
------------------------------------------------------------