NOx-N
Appendix
6 - 3
ba43110e05
01/2011
6.3
RS commands
The RS commands for the TresCon NOx-N all begin with TX where X
is the internal number of the measuring module ( 1 ... 3).
TXCCALI
AutoCal is started. TresCon subsequently goes into the measuring
mode. If the AutoClean interval has expired, AutoClean is called up
first.
Instrument response:
TXCLEAN
AutoClean is started followed by AutoCal. TresCon subsequently goes
into the measuring mode.
Instrument response:
TXCMRUN
TresCon is put into the measuring mode. If the intervals for AutoClean
and AutoCal have expired, the respective routine begins automatically.
Instrument response:
TresCon>
TXCSTOP
TresCon is put into the Stop mode.
Instrument response:
TresCon>
TresCon>
--------------------------------------------
NOx-N Module 1 : Start AutoCal
22.04.03 15:21:49
--------------------------------------------
TresCon>
TresCon>
--------------------------------------------
NOx-N Module 1 : Start AutoClean
22.04.03 15:21:49
--------------------------------------------
TresCon>