13 User-defined methods
Data format of the method data:
The data of user-defined methods is transferred in a
string. The individual datablocks of the string are sep-
arated by spaces:
Example: CEME 301 Test1 690nm mmol/l
C6H5OH
0.009 2.12 0.1 22.3 14 0.1 10.0 1.0
<CR>
Error message "INVALID COMMAND"
The following inputs lead to this error message:
●
Slope = 0 or < -32000, > 32000
●
Zero point > 32000
●
Meas. range begin < 0 or > 32000
●
Meas. range end < 0 or > 32000
●
Meas. range begin > Meas. range end
●
Reference cell not equal to 10 mm, 14 mm, 20 mm
or 50 mm
●
Designation of the wavelength does not agree with
the photometer
●
Resolution not 0.1, 0.01, ...
Further information is given in the chapter
.
Datablock
Maximum no. of characters
Example
Input user-defined methods
4
CEME
Method number
3
301
Method designation
5
TEST1
Wavelength
5
690nm
Unit
9
mmol/l
Citation form
12
C6H5OH
Zero point
5
0.009
Slope
5
2.12
Meas. range begin
5
0.1
Meas. range end
5
22.3
Reference cell
2
14 (= round)
Resolution
5
0.1
AQA theor. value
10
10.0
AQA tolerance
10
1.0
13.2
Entering characteristics via the PC
Содержание PhotoLab Spektral
Страница 2: ......
Страница 31: ...Edition date 06 00 Order no ba 41111e04 Functional description Edition date 04 2004 Order no ba41111e08...
Страница 32: ......
Страница 38: ...8 List of contents...
Страница 128: ...98 13 User defined methods...
Страница 132: ...102 14 Photometer setup...
Страница 140: ...110 16 RS 232 C interface...
Страница 144: ...114 18 Exchanging the Memory Card...
Страница 152: ...122 21 Appendix 21 1 Menu structure of the setup menu...
Страница 153: ...123 21 Appendix...
Страница 154: ...124 21 Appendix...
Страница 160: ......
Страница 161: ......