Continuing for two-point
calibration
(Cal. type TEC)
8
Thoroughly rinse the electrode with distilled water.
9
Immerse the electrode in buffer solution 2.
10
If the Set temperature menu item appears, measure and enter
the temperature of the buffer manually (measurement without
temperature sensor).
11
Using <
▲
> <
▼
>, select Continue and press <START/EN-
TER>. The buffer is measured.
The measured value is checked for stability (AutoRead).
12
Wait for the end of the AutoRead measurement or Terminate
AutoRead with <START/ENTER> and take over the calibration
value.
The calibration display for the next buffer appears.
13
For two-point calibration, select Exit with 2 point calibration with
<
▲
> <
▼
> and confirm with <START/ENTER>.
The calibration is completed as a two-point calibration.
The new calibration values are displayed as an informative
message (
i
).
You have the following options:
z
Accept the new calibration values with <START/ENTER>.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
z
To switch to the measured value display without accepting
the new calibration values, press <M> (short pressure) or
<ESC>.
i
Buffer value = 10.01
i
U = -177 mV
i
Temperature =
24.8 °C
Terminate AutoRead
pH & ORP \ Calibration
i
Buffer recognition TEC
i
Immerse sensor in buffer 3
Exit with 2 point calibration
Continue
pH & ORP \ Calibration
Содержание pHotoFlex
Страница 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Страница 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Страница 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Страница 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Страница 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Страница 65: ...pHotoFlex pHotoFlex Turb Operation 65 ba75491e05 04 2009 9 Repeat the steps 2 to 8 for further samples...
Страница 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Страница 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Страница 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...