pHotoFlex / pHotoFlex Turb
Operation
53
ba75491e05
04/2009
4.6.3
Measuring the ORP voltage
Note
ORP electrodes are not calibrated. However, you can check ORP elec-
trodes using a test solution.
AutoRead
(drift control)
The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured pa-
rameter flashes until a stable measured value is available.
Criteria
With identical measurement conditions, the following applies:
4.6.4
Settings for pH and ORP measurements
Overview
For pH and ORP measurements, the following settings are available in
the Configuration / pH & ORP menu:
z
Measured parameter
z
Calibration record (display, print)
z
Selecting the calibration type
z
Entering the Calibration interval
z
Selecting the Unit for slope
z
Selecting the Temperature unit
z
Reset
1
Perform the preparatory activities according to section 4.6.1.
2
Submerse the ORP electrode in the sample.
3
Press the <M> key (short pressure) repeatedly until the mea-
sured parameter, ORP is selected.
pH & ORP \ ORP
157
mV
25.0
°C
[AR]
01.02.05 15:12
Measured
parameter
Reproducibility
Response time
ORP voltage
better than 1 mV
> 30 seconds
Содержание pHotoFlex
Страница 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Страница 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Страница 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Страница 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Страница 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Страница 65: ...pHotoFlex pHotoFlex Turb Operation 65 ba75491e05 04 2009 9 Repeat the steps 2 to 8 for further samples...
Страница 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Страница 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Страница 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...