Zero adjustment
The zero adjustment, i. e. measuring and storing the absorbance of a
cell filled with water, is necessary after the meter is switched on.
Additionally, we recommend to carry out a zero adjustment if the ambi-
ent temperature has changed.
Only perform the zero adjustment against distilled water in an optically
perfect cell. The zero adjustment must be performed separately for
each cell type.
5
Clean the cell (see section 5.2.2).
6
Insert the cell (see section 4.2).
7
Start the measurement with <START/ENTER>.
The measurement result is displayed when the measurement
is finished.
1
Press the <M> key (long pressure) repeatedly until the Pho-
tometry measuring mode is selected.
2
Press the <M> key (short pressure) repeatedly until the mea-
sured variable, Concentration is selected.
3
Press the <CAL/ZERO> key.
The menu for adjustment measurements opens up.
Photometry \ Absorbance
i
Insert sample
i
Start measurement
with <START>
16 mm
610 nm
01.02.05 15:12
Photometry \ Absorbance
0.532
16 mm
610 nm
01.02.05 15:12
Содержание pHotoFlex
Страница 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Страница 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Страница 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Страница 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Страница 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Страница 65: ...pHotoFlex pHotoFlex Turb Operation 65 ba75491e05 04 2009 9 Repeat the steps 2 to 8 for further samples...
Страница 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Страница 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Страница 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...