pH/ION 7320
Ion concentration
ba75935e04
12/2017
61
8.3.4
Sample subtraction
With the "Sample subtraction" procedure, a known amount of sample is added
to the standard solution (complexing agent or precipitating agent).
The ion concentration in the sample is calculated from the change in potential.
The sample subtraction is one method for the indirect determination of ions. It
is used for the determination of ions for which no direct determination is possi-
ble.
Calibration
1.
Select the measuring method (see section 8.3 S
2.
If necessary, set the ion type for the ion in the standard (
Ion type (sub)
)
with
<
> <
>
and
<MENU/ENTER>
.
On selection of the ion type that can be defined (ION):
Using
<
> <
>
and
<MENU/ENTER>
, set the valence (
Valency
) and
molar weight (
Molar mass
) for the ion in the standard solution.
3.
Select
Continue
with
<
> <
>
and confirm with
<MENU/ENTER>
.
i
Set the ion type for the standard.
Ion type
S
Continue
Sample subtraction
12.12.2012
12:12
Back
4.
Prepare two standard calibration solutions.
5.
Perform a two-point calibration according to user guidance.
6.
The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.
CALIBRATION
12.12.2012 12:12:12
#1 0.010 mg/l
79.2 mV 25.0 °C
#2 0.100 mg/l
20.2 mV 25.0 °C
Slp.: -59.0 mV
++ (Ion type: S)
12.12.2012
12:12
Continue
USB output
Содержание pH/ION 7320
Страница 2: ...2 ba75935e04 12 2017 pH ION 7320 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 6: ...Contents pH ION 7320 6 ba75935e04 12 2017 ...
Страница 86: ...Transmitting data USB interface pH ION 7320 86 ba75935e04 12 2017 ...
Страница 92: ...What to do if pH ION 7320 92 ba75935e04 12 2017 ...
Страница 103: ......