Ion concentration
pH/ION 7320
48
ba75935e04
12/2017
8.2
Calibration
8.2.1 Why calibrate?
Ion-selective electrodes age and are temperature-dependent. This changes
the slope. As a result, an inexact measured value is displayed. Calibration
determines the calibration line of the electrode and stores this value in the
meter.
Thus, you should calibrate before each measurement and at regular intervals.
8.2.2 When to calibrate?
Before any measurement if possible
After connecting another ISE electrode
When the sensor symbol flashes, e.g. after a voltage interruption (empty
batteries)
8.2.3 Calibration (ISE Cal)
ISE Cal is the conventional
two-point
to
seven-point calibration procedure
that uses 2 to 7 freely selectable standard solutions. The concentration
expected for the measurement determines the concentration of the calibration
standards.
1.
Connect the ISE combination electrode to the meter.
The pH/mV/ISE measuring window is displayed.
2.
Keep the standard solutions ready.
3.
If necessary, measure the temperature of the standard solutions with a
thermometer.
4.
If necessary, change the unit of the measurement result and calibration
standards in the
ISE setup
/
Unit
menu.
5.
In the measured value display, select the ISE measuring window with
<
> <
>
and
<M>
.
6.
Start the calibration with
<CAL>
.
The calibration display appears.
ISE
0.0
24.8 °C
Ag
0.010
mg/l
1
12.12.2012
12:12
Standard
mV
Содержание pH/ION 7320
Страница 2: ...2 ba75935e04 12 2017 pH ION 7320 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 6: ...Contents pH ION 7320 6 ba75935e04 12 2017 ...
Страница 86: ...Transmitting data USB interface pH ION 7320 86 ba75935e04 12 2017 ...
Страница 92: ...What to do if pH ION 7320 92 ba75935e04 12 2017 ...
Страница 103: ......