Operation
pH 3300i/3400i
18
ba75682e02
03/2008
4.3
Calibrating
Why calibrate?
pH electrodes age. This changes the asymmetry (zero point) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines the current values of the asymmetry and slope of the electrode
and stores them in the measuring instrument.
Thus, you should calibrate at regular intervals.
When to calibrate?
z
After connecting another electrode
z
When the sensor symbol flashes:
– after the calibration interval has expired
– after a voltage interruption, e. g. after changing the batteries
Calibration interval
The calibration interval (Int 3) is set to 7 days in the factory. You can select
the interval in the range of 1 ... 999 days (see section 4.6 C
You can select between 3 calibration procedures:
AutoCal TEC
is specially adapted to the WTW technical buffer solutions as a fully automat-
ic
two-point
or
three-point calibration
. The buffer solutions are automati-
cally recognized by the measuring instrument. Depending on the instrument
setting (see section 4.6 C
), the instrument displays the relevant
buffer nominal value or the current electrode voltage in mV. The calibration
can be terminated after the first buffer solution. This corresponds to a
single-
point calibration
. For this, the instrument uses the Nernst slope (-59.2 mV/
pH at 25 °C) and determines the asymmetry of the electrode.
AutoCal DIN
is specially adapted to the permanently programmed buffer solutions in ac-
cordance with DIN19266 as a fully automatic
two-point
or
three-point cali-
bration
. The buffer solutions are automatically recognized by the measuring
instrument. Depending on the instrument setting (see section
4.6 C
), the instrument displays the relevant buffer nominal val-
ue or the current electrode voltage in mV. The calibration can be terminated
after the first buffer solution. This corresponds to a
single-point calibration
.
For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and de-
termines the asymmetry of the electrode.
ConCal
This function is a conventional
two-point calibration
using two buffer solu-
tions (pH 7.0 ± 0.5 and any other buffer solution) or a
single-point calibra-
tion
using any buffer solution and is used as a high-speed method.
AutoRead
When calibrating with AutoCal TEC and AutoCal DIN, the
AutoRead
function
is automatically activated.
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing
<RUN/ENTER>
.
Displaying
calibration data
Each time the instrument is switched on, the calibration data are shown on
the display for a short time (see section 4.1 S
-
Содержание pH 3300i
Страница 4: ...List of contents pH 3300i 3400i 4 ba75682e02 03 2008 7 Technical data 57 8 Lists 61...
Страница 31: ...pH 3300i 3400i Operation 31 ba75682e02 03 2008 15 To return to the measuring mode Press the M key...
Страница 56: ...What to do if pH 3300i 3400i 56 ba75682e02 03 2008...
Страница 60: ...Technical data pH 3300i 3400i 60 ba75682e02 03 2008...
Страница 66: ...Lists pH 3300i 3400i 66 ba75682e02 03 2008...
Страница 68: ...Index pH 3300i 3400i 68 ba75682e02 03 2008...