OxiTop
®
Control OC 110
Operating mode: BOD Standard / BOD Routine
ba31116e05
07/2014
37
5.5.2
Excluding a curve (
BOD Standard
operating mode)
This function is used to temporarily exclude a single curve (e.g. an
outlier) from the evaluation and averaging of an overall sample.
The function "List of measuring heads“ (see section 5.4
SAMPLE STATUS TYPE
), page 28) is used
to find leaky or defective single samples.
The curve is only excluded temporarily!
The data stock of the
SAMPLE STATUS TYPE
list (
sample management) does not
change. The excluded curve is present again when the call is
repeated.
Display of single curves:
Select curve.
Use
<
><
>
to select
-
Exclude curve
.
- Exclude curve
is preselected.
- Back
: Returns to the previous
menu.
Message
Curve
excluded!
.
Updated display (curve
excluded, mean value recalcu-
lated).
Use
<PRINT>
to print the
updated results with curves
(without the excluded curve).
RUN/ENTER
RUN/ENTER
RUN/ENTER
Содержание OxiTop Control OC 110
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Страница 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Страница 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Страница 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Страница 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Страница 101: ......