Dissolved oxygen
Oxi 7310
28
ba75931e03
12/2011
5.2.6
Calibration data
Displaying the cali-
bration data
The calibration record of the last calibration is to be found under the menu item,
<F1>
/ [Menu] /
Calibration
/
Calibration record
. To open it, press the
<CAL_>
key in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with
<F2>
[USB output]
.
Displaying the
cali-
bration data storage
The calibration records of the last calibrations (up to 10) are available in the
menu,
<F1>
/[Menu] /
Calibration
/
Calibration data storage
and in the menu,
<F1_>
/[Menu] /
Storage & config
/
Data storage
/
Calibration data storage
.
Calibration evalua-
tion
After the calibration, the meter automatically evaluates the current status of the
calibration. The evaluation appears on the display and in the calibration record.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/func-
tion
Description
Calibration
/
Calibration data stor-
age
/
Display
or
Data storage
/
Calibration data stor-
age
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Quit the display with
<F1>
/
[Back]
or
<ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
or
Data storage
/
Calibration data stor-
age
/
Output to USB
-
Outputs the calibration records to
the interface.
Содержание Oxi 7310
Страница 1: ...ba75931e03 12 2011 Oxi 7310 Oxi 7310P Operating manual D O meter Oxi 7310P Oxi 7310 Printermodul ...
Страница 56: ...Glossary Oxi 7310 56 ba75931e03 12 2011 ...
Страница 58: ...Index Oxi 7310 58 ba75931e03 12 2011 ...
Страница 59: ......