Oxi 1970i
Resetting to default settings
27
ba75676e03
05/2012
Resetting to default settings
You can reset the measurement parameters and the configuration to the
delivery status separately from one another (initialization).
Measurement
parameters
The following measurement parameters can be reset to the delivery status:
Configuration
parameters
The following configuration parameters (InI) can be reset to the delivery
status:
Reset procedure:
Disposal
At the end of its operational lifetime, the meter must be returned to the
disposal or return system statutory in your country. If you have any questions,
please contact your supplier.
Oxygen (O
2
InI)
Measuring mode
D. O. concentration
AutoRange automatic switchover of
the measurement range
On (YES)
Relative slope
1.00
Correction factor of the relative
slope
1.000
Salinity input
0.0
Salinity correction
Off
Baud rate
4800
Interval 1
(automatically saved)
OFF
Interval 2
(for data transmission, Oxi 340i
only)
OFF
Press <CAL> while pressing the <RUN/ENTER> key.
To reset the measurement parameters, use <
▲
> <
▼
> to select YES and
confirm with <RUN/ENTER>, or
Continue to the configuration parameters without resetting (InI) with the
<RUN/ENTER> key.
After the InI configuration parameters, the instrument switches to the last
active measurement mode.