Configuring
Oxi 1970i
26
ba75676e03
05/2012
Configuring
Note
You can leave the configuration menu at any time with <M>. The parameters
that have already been changed are stored.
Procedure for configuring (factory settings appear in bold typeface):
Switch off the measuring instrument.
Press <ON/OFF>> while pressing the <M> key.
Select the required Baud rate with <
▲
> <
▼
>.
Selection: 1200, 2400, 4800, 9600 Baud.
Then press the <RUN/ENTER> key.
The air pressure is only displayed if an oxygen sensor is connected.
Otherwise, "---" appears.
Continue with the <RUN/ENTER> key.
Select the required calibration interval for oxygen measurements with
<
▲
> <
▼
>.
Selection: 1 ... 14 ... 999 d.
Then press the <RUN/ENTER> key.
Switch the automatic measuring range selection AutoRange on/off with
<
▲
> <
▼
>.
Selection: YES (On), no (Off).
Then press the <RUN/ENTER> key.
Select the date and time step-by-step with <
▲
> <
▼
>. In between, press
the <RUN/ENTER> key each time.
When the last <RUN/ENTER> key has been pressed, the instrument
switches to the last active measuring mode.
Sa
T
°
C
Year
Ident
No.
Day.Month
Time
cm
1/
K
/
%
8 0 0
4
Baud
d
C A
pH
pH
/
Sal
mg/ l
O
TP
°
C
S
Year
Ident
No.
Day.Month
Sal
C
Time
Baud
cm
1/
K
/
%
mV
m
cm
S/
%
9 4 9
P
1
cm
1/
TP
°
C
cm
S/
Sal
%
S
%
m
mg/ l
Year
Ident
No.
Day.Month
Sal
pH
Baud
K
/
pH
/
mV
1 4
3
n t
d
I
Time
O
Tref25
Time
d
I
cm
1/
TP
°
C
cm
S/
Sal
%
S
%
m
mg/ l
Year
Ident
No.
Day.Month
Tref20
Sal
pH
Auto
Store
Baud
K
/
REL2
pH
/
mV
REL1
n
A r
E S
Y
ARng
O
Day.Month
0 9 0 4