Maintenance

17

ba75408e04

12/2005

Maintenance

In the case of increased response times or drifting measured values, 
store the electrode in the standard solution ES/NO3_ISA-50 (ordering 
information, see section A

CCESSORIES

) for several hours. 

Subsequently, recalibrate.

Cleaning

Storage

1 to 2 days

Attach the wet cap to the combined electrode with the exchange 
electrode screwed in. Before doing so, wet the foam padding in the wet 
cap with ES/NO3_ISA-50 standard solution (ordering information, see 
section A

CCESSORIES

).

More than two days

Uninstall the exchange electrode and store the components as follows.

Reference electrode: 
Screw in the blind plug, wet the foam padding in the wet cap with 
deionized water and plug on the wet cap.

Exchange electrode:
Screw the exchange electrode into the nut up to the stop. Before doing 
so, wet the foam padding in the watering cap with ES/NO3_ISA-50 
standard solution (ordering information, see section A

CCESSORIES

). 

Subsequently, plug the exchange electrode into the wet watering cap.

Contamination

Cleaning

Coarse contamination on the 
junction of the NitraLyt

®

NOA 

reference electrode

Carefully brush off contamination 
under running water using a soft 
toothbrush.

Contamination on the 
membrane of the 
NitraLyt

®

NOA/AT exchange 

electrode

Splash the electrode with water. 
Remove coarse contamination using 
a soft toothbrush under running 
water.

Содержание NitraLyt NOA

Страница 1: ...ba75408e04 12 2005 NitraLyt NOA NitraLyt NOA AT Operating Manual NitraLyt nitrate combination electrode NitraLyt NOA reference electrode NitraLyt NOA AT exchange electrode NitraLyt NOA AT NitraLyt NOA...

Страница 2: ...ctrode Also we cannot guarantee that there are absolutely no errors in this manual Therefore we are sure you will understand that we cannot accept any legal claims resulting from the data figures or d...

Страница 3: ...raLyt NOA AT Contents 13 ba75408e04 12 2005 Contents Overview 14 Assembly 15 Response times 16 Aging 16 Factors affecting the measured values 16 Maintenance 17 Cleaning 17 Storage 17 Accessories 18 Te...

Страница 4: ...efficient aeration control The NitraLyt combination electrode is used in the NitraLyt 700 IQ sensor Structure of the NitraLyt combination electrode Caution The membrane and junction of the electrode m...

Страница 5: ...if you want to store the reference electrode 2 Remove the watering cap 7 and nut 8 from the exchange electrode Store both accessories You need them if you want to store the reference electrode 3 Scre...

Страница 6: ...ndergoes a natural aging process The response time increases and the slope decreases with the age of the electrode The following factors shorten the lifetime considerably Incorrect storage Special mea...

Страница 7: ...store the components as follows Reference electrode Screw in the blind plug wet the foam padding in the wet cap with deionized water and plug on the wet cap Exchange electrode Screw the exchange elect...

Страница 8: ...WTW catalog and on the Internet Note Information on the fundamentals of measuring with ion sensitive electrodes are given in the WTW primer ION SELECTIVE MEASUREMENT IN ONLINE ANALYSIS Designation Or...

Страница 9: ...F Response time at 20 C 3 min when concentration changes from 5 mg l to 50 mg l NO3 N Type of protection IP 68 0 2 bar complete combination electrode mounted in the NitraLyt 700 IQ sensor Connection...

Страница 10: ...Wissenschaftlich Technische Werkst tten GmbH Dr Karl Slevogt Stra e 1 D 82362 Weilheim Germany Tel 49 0 881 183 0 49 0 881 183 100 Fax 49 0 881 183 420 E Mail Info WTW com Internet http www WTW com...

Отзывы: