Commissioning
NitraLyt
®
700 IQ
3 - 6
ba75397d03
12/2004
3.4
Carrying out the settings for the sensor on the termi-
nal of the IQ S
ENSOR
N
ET
system
The following settings can be carried out for the sensor:
Menu item
Possible settings
Explanations
Measuring mode
!
NO3
!
NO3-N
!
mV
The citation form of the mass concentra-
tion or the voltage of the electrode.
Measuring range
(
NO3
)
!
AutoRange
!
5 ... 4500 mg/l
!
0.5 ... 450.0 mg/l
Two measuring ranges can be selected.
With
AutoRange
, the instrument automati-
cally switches to the suitable measuring
range.
Measuring range
(
NO3-N
)
!
AutoRange
!
1 ... 1000 mg/l
!
0.1 ... 100.0 mg/l
Two measuring ranges can be selected.
With
AutoRange
, the instrument automati-
cally switches to the suitable measuring
range.
Measuring range
(
mV
)
!
-2000 ... 2000 mV
Fixed range
Temperature mode
!
°C
!
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).